Intuitive Surgical, Inc. recall records

69 reported ongoing

72 records in this archive. Latest source date Apr 24, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

8MM, Tenaculum Forceps REF 470207 Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical tasks including manipu...

Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

Class IIReported ongoing

Distribution: U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NC, ND, NE, N...

Official source for 8MM, Tenaculum Forceps REF 470207 Instrument used with the da Vinci X/Xi Surgical Systems in a wi...
FDA device enforcement

8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical ta...

Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

Class IIReported ongoing

Distribution: U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NC, ND, NE, N...

Official source for 8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surg...
FDA device enforcement

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Conta...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.

Official source for da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380...
FDA device enforcement

Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion...

Class IIReported ongoing

Distribution: US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea

Official source for Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Reasons in these listed records

  • Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
  • Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments
  • Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support...
  • There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.

These examples come from this result page and are not a ranking of every recall reason.

Checking Intuitive Surgical, Inc.

Is there a recall on intuitive surgical, inc. right now?

This archive contains 72 records, including 69 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking