Ion Endoluminal System, REF: 380748-65 — FDA record Z-2720-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ion Endoluminal System, REF: 380748-65
Reason
There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
Classification
Class II
Recalling firm
Intuitive Surgical, Inc.
Quantity
3 units
Agency-reported status
Ongoing
Source record identifier
Z-2720-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 00886874116234/ System Serial Number/System Cart Serial Number: EN1230/10591271; EN1241/10591273; EN1231/10583014

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AZ and CA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 4, 2024
Recall initiation date
Jul 12, 2024
Center classification date
Aug 23, 2024