Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098 — FDA record Z-1518-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 13, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
Reason
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
Classification
Class II
Recalling firm
Intuitive Surgical, Inc.
Quantity
53
Agency-reported status
Ongoing
Source record identifier
Z-1518-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPart/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 17, 2024
Recall initiation date
Mar 13, 2024
Center classification date
Apr 11, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.