da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40 — FDA record Z-0964-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
Reason
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
Classification
Class II
Recalling firm
Intuitive Surgical, Inc.
Quantity
144
Agency-reported status
Ongoing
Source record identifier
Z-0964-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 31, 2025
Recall initiation date
Aug 21, 2024
Center classification date
Dec 19, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.