Recall archive / FDA device enforcement
da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40 — FDA record Z-0964-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
- Reason
- Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
- Classification
- Class II
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 144
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0964-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Dec 31, 2025
- Recall initiation date
- Aug 21, 2024
- Center classification date
- Dec 19, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.