Records and reported status

FDA device enforcement

Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator System, Part Number 6688600 Servo-u MR ventilator: Part Number 6888800 Servo-ai...

Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to hypoventilation and consequently desatur...

Class IReported ongoing

Distribution: Worldwide distribution. US nationwide, Poland, Spain, France, Germany, Czech Republic, Italy, Belgium, Estonia, Greece, Finland, Portugal, Netherlands, Romania, Austria, Slovaki...

Official source for Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator...
FDA device enforcement

VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961 VLT600DF AIM STP ARDVST229000A VST60SF AIM ARD...

Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the patient body or in the...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, GA, IA, ID, KS, KY, LA, MA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, VA, WA.

Official source for VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811...
FDA device enforcement

Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700

Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.

Class IReported terminated

Distribution: US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri, and Virginia

Official source for Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the ent...
FDA device enforcement

Flow-e Anesthesia System : Intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887900

Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.

Class IReported terminated

Distribution: US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri, and Virginia

Official source for Flow-e Anesthesia System : Intended for use in administering anesthesia while controlling the ent...
FDA device enforcement

Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, etc. Model: CM...

Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.

Official source for Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfect...
FDA device enforcement

Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD

Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.

Official source for Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for th...
FDA device enforcement

Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.

Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.

Class IIReported ongoing

Distribution: Worldwide distribution: (US) AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, SD, TX, VA, WA, WV. (OUS) Austra...

Official source for Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.
FDA device enforcement

Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700 07325710009765 Flow-e Anesth...

A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.

Class IReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Ir...

Official source for Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Produc...

Reasons in these listed records

  • Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to hypoventilation and consequently desaturation, with patient ...
  • Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the patient body or in the operating field.
  • Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
  • Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

These examples come from this result page and are not a ranking of every recall reason.

Checking Getinge Usa Sales Inc

Is there a recall on getinge usa sales inc right now?

This archive contains 48 records, including 46 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking