Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of p... — FDA record Z-1278-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 2, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700
Reason
Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
Classification
Class I
Recalling firm
Getinge Usa Sales Inc
Quantity
17 units US
Agency-reported status
Terminated
Source record identifier
Z-1278-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 07325710009765 Serial Numbers: 4003, 4777, 4783, 5024, 5025, 5026, 5027, 5028, 5043, 5044, 5045, 5046, 5047, 5048, 5049, 5050, 5078. (Note: UDI was corrected July 7, 2022)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri, and Virginia

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 29, 2022
Recall initiation date
May 2, 2022
Center classification date
Jun 23, 2022
Termination date
Aug 6, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.