Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: F... — FDA record Z-0468-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 9, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700 07325710009765 Flow-e Anesthesia System 6887900 07325710010457 Flow-i Anesthesia System C20 6888520 07325710010617 Flow-i Anesthesia System C30 6888530 07325710010624 Flow-i Anesthesia System C40 6888540 07325710010631 Flow-i C20 Anesthesia System 6677200 07325710001349 Flow-i C30 Anesthesia System 6677300 07325710001349 Flow-i C40 Anesthesia System 6677400 07325710001318
Reason
A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
Classification
Class I
Recalling firm
Getinge Usa Sales Inc
Quantity
50 US; 898 OUS
Agency-reported status
Ongoing
Source record identifier
Z-0468-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 07325710008430 Serial Nos. 15580 15890 15924 16284 16637 16837 16970 15582 15891 15925 16285 16638 16840 16971 15721 15894 16268 16345 16639 16842 15722 15914 16269 16352 16640 16843 15725 15917 16270 16354 16690 16844 15726 15919 16271 16632 16828 16961 15728 15920 16272 16633 16832 16964 15729 15921 16273 16634 16833 16965 15731 15922 16276 16635 16834 16966 15732 15923 16283 16636 16836 16968

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Iraq, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Morocco, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 19, 2022
Recall initiation date
Dec 9, 2021
Center classification date
Jan 10, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.