Getinge Usa Sales Inc recall records

46 reported ongoing

48 records in this archive. Latest source date Jan 15, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Model Numb...

Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the followi...

Class IIReported ongoing

Distribution: US Nationwide distribution.

Official source for Meera Mobile Operating Table-to support and position the patient immediately before, during and a...
FDA device enforcement

Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400

Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

Class IReported ongoing

Distribution: US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.

Official source for Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-...
FDA device enforcement

Getinge Flow-c Anesthesia System, Part No. 6887700

Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

Class IReported ongoing

Distribution: US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.

Official source for Getinge Flow-c Anesthesia System, Part No. 6887700
FDA device enforcement

Getinge Flow-e Anesthesia System, Part No. 6887900

Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

Class IReported ongoing

Distribution: US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.

Official source for Getinge Flow-e Anesthesia System, Part No. 6887900
FDA device enforcement

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinicia...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.

Official source for Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
FDA device enforcement

Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN. The countries of Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Bulgaria (BG), Canada (CA),...

Official source for Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
FDA device enforcement

Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support, monitoring and treatment of pediatric and adult patients. Model Number: 6882000...

Device does not meet regulatory requirements of stability while stationary, which requires mechanical equipment, other than fixed mechanical equipment, that is intended to be used on the floor or on a table shall be permanently marked with a clearly legible...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Australia (AU), Brazil (BR), China (CN), Egypt (EG), EEA, Hong Kong (HK), Israel (IL), Japan (JP), Mexico (MX), Pakis...

Official source for Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support,...

Reasons in these listed records

  • A potential for a light system to fall in the operating room.
  • Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably forese...
  • Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
  • The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emer...

These examples come from this result page and are not a ranking of every recall reason.

Checking Getinge Usa Sales Inc

Is there a recall on getinge usa sales inc right now?

This archive contains 48 records, including 46 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking