Recall archive / FDA device enforcement
Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN — FDA record Z-0239-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
- Reason
- The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.
- Classification
- Class II
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 5 US; 1008 OUS
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0239-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | No UDI - product discontinued in 2008. Part Number 1140.20F0: Serial Nos. 3, 8, 9, 10 Part Number 1140,20AN: Serial No. 851 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN. The countries of Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Bulgaria (BG), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Estonia (EE), Finland (FI), France (FR), Germany (DE), Greece (GR), Hungary (HU), Indonesia (ID), Israel (IL), Italy (IT), Japan (JP), Jordan (JO), Kuwait (KW), Lithuania (LT), Luxembourg (LU), Mexico (MX), Montenegro (ME), Morocco (MA), Netherlands (NL), Norway (NO), Pakistan (PK), Peru (PE), Poland (PL), Portugal (PT), Romania (RO), Russia (RU), Saudi Arabia (SA), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), Tunisia (TN), Turkey (TR), United Arab Emirates (AE), United Kingdom (GB), Yemen (YE).
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 23, 2022
- Recall initiation date
- Oct 14, 2022
- Center classification date
- Nov 15, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.