Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900) — FDA record Z-0422-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 7, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
Reason
The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.
Classification
Class II
Recalling firm
Getinge Usa Sales Inc
Quantity
65 US
Agency-reported status
Ongoing
Source record identifier
Z-0422-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityFlow-c UDI-DI: 7325710010457 Lot/Serial Numbers 4003 4777 4783 5024 5025 5026 5027 5028 5043 5044 5045 5046 5047 5048 5049 5050 5078 Flow-e UDI-DI: 7325710009765 Lot/Serial Numbers 50012 50013 50014 50015 50016 50017 50030 50031 50033 50034 50035 50036 50037 50038 50039 50040 50041 50042 50043 50046 50047 50048 50051 50052 50053 50054 50055 50056 50057 50058 50059 50060 50061 50062 50069 50070 50071 50072 50073 50074 50084 50085 50086 50087 50088 50089 50090 50091 50092 50093 50094 50096 50097 50108 50109 50110 50111 50132

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Nov 7, 2022
Center classification date
Dec 7, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records