Shimadzu Medical Systems recall records

6 reported ongoing

6 records in this archive. Latest source date Apr 20, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 /...

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ,...

Official source for Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST saf...
FDA device enforcement

MODEL: X-RAY R/F SYSTEM FLUOROspeed X1

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Class IIReported ongoing

Distribution: U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A

Official source for MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
FDA device enforcement

MODEL: X-RAY TV SYSTEM SONIALVISION G4

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Class IIReported ongoing

Distribution: U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A

Official source for MODEL: X-RAY TV SYSTEM SONIALVISION G4
FDA device enforcement

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Class IIReported ongoing

Distribution: U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A

Official source for MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
FDA device enforcement

Digital Angiography System Bransist safire Generator Model D150GC-40

Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.

Class IIReported ongoing

Distribution: U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.

Official source for Digital Angiography System Bransist safire Generator Model D150GC-40
FDA device enforcement

Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40

Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.

Class IIReported ongoing

Distribution: U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.

Official source for Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40

Reasons in these listed records

  • Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.
  • It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
  • Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.

These examples come from this result page and are not a ranking of every recall reason.

Checking Shimadzu Medical Systems

Is there a recall on shimadzu medical systems right now?

This archive contains 6 records, including 6 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking