Digital Angiography System Bransist safire Generator Model D150GC-40 — FDA record Z-0654-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 16, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Digital Angiography System Bransist safire Generator Model D150GC-40
- Reason
- Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
- Classification
- Class II
- Recalling firm
- Shimadzu Medical Systems
- Quantity
- 18
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0654-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Generator Model: D150GC-40 Serial Number Range - 0161Q81103 to 3M99D3A55001 Specific Generator Serial Number/SMS ID: 3M99D3A24001/SMS20120038 3M99D3A4B001/SMS20140084 3M99D3A55001/SMS20150030 0161Q83701/SMS20080542 0161Q82204/SMS20080045 0161Q82302/SMS20080043 0161Q82304/SMS20080046 0261Q80901/SMS20100011 3M7A1600B001/SMS20080286 3M99D2622001/SMSMA110079 0161Q84501/SMS20090038 3M7A1600C001/SMS20200285 3M7A16A1A001/SMSMA110013 3M99D3A1C001/SMS20110202 0161Q84103/SMS20090099 0161Q83603/SMS20080505 0161Q83702/SMS20080580 0161Q83703/SMS20080581 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Mar 9, 2022
- Recall initiation date
- Dec 16, 2021
- Center classification date
- Feb 27, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.