Digital Angiography System Bransist safire Generator Model D150GC-40 — FDA record Z-0654-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 16, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Digital Angiography System Bransist safire Generator Model D150GC-40
Reason
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Classification
Class II
Recalling firm
Shimadzu Medical Systems
Quantity
18
Agency-reported status
Ongoing
Source record identifier
Z-0654-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGenerator Model: D150GC-40 Serial Number Range - 0161Q81103 to 3M99D3A55001 Specific Generator Serial Number/SMS ID: 3M99D3A24001/SMS20120038 3M99D3A4B001/SMS20140084 3M99D3A55001/SMS20150030 0161Q83701/SMS20080542 0161Q82204/SMS20080045 0161Q82302/SMS20080043 0161Q82304/SMS20080046 0261Q80901/SMS20100011 3M7A1600B001/SMS20080286 3M99D2622001/SMSMA110079 0161Q84501/SMS20090038 3M7A1600C001/SMS20200285 3M7A16A1A001/SMSMA110013 3M99D3A1C001/SMS20110202 0161Q84103/SMS20090099 0161Q83603/SMS20080505 0161Q83702/SMS20080580 0161Q83703/SMS20080581

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 9, 2022
Recall initiation date
Dec 16, 2021
Center classification date
Feb 27, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.