Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40 — FDA record Z-0655-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 16, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40
Reason
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Classification
Class II
Recalling firm
Shimadzu Medical Systems
Quantity
28 systems
Agency-reported status
Ongoing
Source record identifier
Z-0655-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGenerator Model: D150GC-40 Serial Number Range - 3M99D44446001to LM99D44B8001 Serial Number/SMS ID: 41E58C765001/SMS20160146 41EFCD78B001/SMS20190007 41EFD3B9A001/SMS20200009 41F005CB8001/SMS20210266 41E58C766001/SMS20160150 41B247746001/SMS20140089 41E58C757001/SMSHP150024 41E58C769001/SMS20160151 41E58C777001/SMS20170121 41EFCD77C001/SMS20180143 41F005B8C001/SMS20190052 41EFCD79A001/SMS20190366 41EFCD79C001/SMS20200285 41EFC73A5001/SMS20200263 41EFCD7A6001/SMS20200267 41EFC73A9001/SMS20200443 41E5E23AA001/SMS20200369 41E58C773001/SMSHP170001 41E58C977001/SMS20170424 41EFFF79A001/SMS20190376 41EFFF79C001/SMS20200031 41EFD3B9C001/SMS20200066 41E58C76B001/SMS20170288 41E58C775001/SMS20170134 41E58C776001/SMS20170135 41F005B91001/SMS20190019 41B24774C001/SMSMA140084 41EFCD7AC001/#N/A

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 9, 2022
Recall initiation date
Dec 16, 2021
Center classification date
Feb 27, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.