When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the ...
Class IIReported ongoing
Distribution: Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, ...
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct...
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by u...
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation...
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibrat...
These examples come from this result page and are not a ranking of every recall reason.
Checking PTW-FREIBURG
Is there a recall on ptw-freiburg right now?
This archive contains 3 records, including 3 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.
How do I check my package or product?
Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.
Does a matching name mean every item is included?
A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.
What if the search finds no matching record?
This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.
Where can I find older records?
Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.