BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is perf... — FDA record Z-2577-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)
Reason
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.
Classification
Class II
Recalling firm
PTW-FREIBURG
Quantity
42 US
Agency-reported status
Ongoing
Source record identifier
Z-2577-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: EPTWS0800320 BeamAdjust All software Versions

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 20, 2023
Recall initiation date
Jun 7, 2023
Center classification date
Sep 11, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records