Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of me... — FDA record Z-1181-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 8, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.
Reason
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
Classification
Class II
Recalling firm
PTW-FREIBURG
Quantity
1 US; 65 worldwide
Agency-reported status
Ongoing
Source record identifier
Z-1181-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: EPTWS070031 Software VERIQA (S070031), version 2.0 and 2.1

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

GA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 28, 2024
Recall initiation date
Feb 8, 2024
Center classification date
Feb 22, 2024