Allura Xper FD10 and FD20. X-ray Imaging System.

The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finla...

Official source for Allura Xper FD10 and FD20. X-ray Imaging System.
FDA device enforcement

Azurion 7 M20. X-ray Image guided Therapy System.

The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finla...

Official source for Azurion 7 M20. X-ray Image guided Therapy System.
FDA device enforcement

Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222...

Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

Class IIReported ongoing

Distribution: Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire...

Official source for Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/...
FDA device enforcement

Zenition 50, Model #718096

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not conn...

Class IIReported ongoing

Distribution: Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Official source for Zenition 50, Model #718096
FDA device enforcement

Zenition 70, Model #718133

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not conn...

Class IIReported ongoing

Distribution: Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Official source for Zenition 70, Model #718133
FDA device enforcement

Veradius Unity, Model #718132

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not conn...

Class IIReported ongoing

Distribution: Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Official source for Veradius Unity, Model #718132

Reasons in these listed records

  • The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
  • Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
  • There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
  • Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.

These examples come from this result page and are not a ranking of every recall reason.

Checking PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Is there a recall on philips medical systems nederland b.v. right now?

This archive contains 208 records, including 208 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking