Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Az... — FDA record Z-1647-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224
Reason
Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
Classification
Class II
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
USA: 1063 R.O.W. 2136
Agency-reported status
Ongoing
Source record identifier
Z-1647-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: System product name Model number UDI (1)Azurion 3M12: 722063 (01)00884838085275(21) 722221 (01)00884838099203(21); (2) Azurion 3M15: 722064 (01)00884838085282(21) 722222 (01)00884838099210(21); (3) Azurion SM12: 722227 (01)00884838099227(21); (4) Azurion SM20: 722228 (01)00884838099234(21); (5) Azurion 7B12/12: 722067 (01)00884838085350(21) 722225 (01)00884838099265(21); (6) Azurion 7820/15: 722068 (01)00884838085367(21) 722226 (01)00884838099272(21); (7) Azurion 7M12: 722078 (01)00884838085251(21) 722223 (01)00884838099241(21); (8) Azurion 7M20: 722079 (01)00884838085268(21) 722224 (01)00884838099258(21), Software: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 7, 2023
Recall initiation date
Mar 21, 2023
Center classification date
May 26, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.