Recall archive / FDA device enforcement
Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Az... — FDA record Z-1647-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224
- Reason
- Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
- Classification
- Class II
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Quantity
- USA: 1063 R.O.W. 2136
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1647-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: System product name Model number UDI (1)Azurion 3M12: 722063 (01)00884838085275(21) 722221 (01)00884838099203(21); (2) Azurion 3M15: 722064 (01)00884838085282(21) 722222 (01)00884838099210(21); (3) Azurion SM12: 722227 (01)00884838099227(21); (4) Azurion SM20: 722228 (01)00884838099234(21); (5) Azurion 7B12/12: 722067 (01)00884838085350(21) 722225 (01)00884838099265(21); (6) Azurion 7820/15: 722068 (01)00884838085367(21) 722226 (01)00884838099272(21); (7) Azurion 7M12: 722078 (01)00884838085251(21) 722223 (01)00884838099241(21); (8) Azurion 7M20: 722079 (01)00884838085268(21) 722224 (01)00884838099258(21), Software: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 7, 2023
- Recall initiation date
- Mar 21, 2023
- Center classification date
- May 26, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.