Zenition 70, Model #718133 — FDA record Z-0648-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 2, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Zenition 70, Model #718133
Reason
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Classification
Class II
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
95 US; 284 ROW
Agency-reported status
Ongoing
Source record identifier
Z-0648-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityWireless Foot switch 3P : 459801238191 Wireless Foot switch 3P: 459801238231

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 23, 2022
Recall initiation date
Dec 2, 2021
Center classification date
Feb 17, 2022