Ossur H / F recall records

7 reported ongoing

7 records in this archive. Latest source date Aug 20, 2024.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101

Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of ...

Class IIReported ongoing

Distribution: US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR

Official source for Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101
FDA device enforcement

RKN130002, RHEO KNEE 2 YR WARRANTY, MD, CE, External lower limb prosthetic component

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, U...

Official source for RKN130002, RHEO KNEE 2 YR WARRANTY, MD, CE, External lower limb prosthetic component
FDA device enforcement

RKN130003, RHEO KNEE 3 YR WARRANTY, MD, CE, External lower limb prosthetic component

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, U...

Official source for RKN130003, RHEO KNEE 3 YR WARRANTY, MD, CE, External lower limb prosthetic component
FDA device enforcement

RKNXC0003, RHEO KNEE XC 3 YR WARRANTY, MD, CE, External lower limb prosthetic component

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, U...

Official source for RKNXC0003, RHEO KNEE XC 3 YR WARRANTY, MD, CE, External lower limb prosthetic component
FDA device enforcement

RKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, U...

Official source for RKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component
FDA device enforcement

Power Knee, REF: PKA10001, and PKA10003; with Power Knee Battery, REF: PKA10002

Knee batteries may need replacement with conforming batteries to ensure there isn't a failure to turn on the device, or an unexpected shutdown, which is a momentary loss of power to the knee where the user may experience loss of support that can lead to a f...

Class IIReported ongoing

Distribution: US Nationwide distribution including in the states of MA, SC, MD, TN, ND, VA, IL, NJ, FL, MI, NY, AL, TX, OK, IN, GA, CA, PA, NM, VT, NV, MS, HI, OH, WI, LA, IA, NE, MN, AR, OR,...

Official source for Power Knee, REF: PKA10001, and PKA10003; with Power Knee Battery, REF: PKA10002
FDA device enforcement

Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body,...

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of...

Class IIReported ongoing

Distribution: US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR

Official source for Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, RE...

Reasons in these listed records

  • Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
  • Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
  • Knee batteries may need replacement with conforming batteries to ensure there isn't a failure to turn on the device, or an unexpected shutdown, which is a momentary loss of power to the knee where the user may experience loss of support that can lead to a fall and potential in...
  • A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosth...

These examples come from this result page and are not a ranking of every recall reason.

Checking Ossur H / F

Is there a recall on ossur h / f right now?

This archive contains 7 records, including 7 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking