Recall archive / FDA device enforcement
Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101 — FDA record Z-0006-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 20, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101
- Reason
- Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
- Classification
- Class II
- Recalling firm
- Ossur H / F
- Quantity
- 21,020
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0006-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI/GTIN/Lot Range/Sold in the Period, Date Range: MJS-101/05690967817008/MX220516 to MX230104/2022-05-19 to 2023-06-05, MJSR-101/05690967817145/MX220516 to MX230104/2022-05-19 to 2023-06-05 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Florida
- Georgia
- Iowa
- Idaho
- Illinois
- Indiana
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Michigan
- Minnesota
- Missouri
- Montana
- North Carolina
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Washington
Dates in the source
- Report date
- Oct 9, 2024
- Recall initiation date
- Aug 20, 2024
- Center classification date
- Oct 1, 2024