Recall archive / FDA device enforcement
Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock... — FDA record Z-1906-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020
- Reason
- A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.
- Classification
- Class II
- Recalling firm
- Ossur H / F
- Quantity
- 4,190
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1906-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- Nebraska
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 21, 2023
- Recall initiation date
- May 12, 2023
- Center classification date
- Jun 13, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.