Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recall records

130 reported ongoing

134 records in this archive. Latest source date May 10, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers: DVPA2D1, DVPA2D4; b. Cobalt XT DR ICD, Model Numbers: DDPA2D1, DDPA2D4

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers: DVPA2D1, DVPA...
FDA device enforcement

Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D...

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a....
FDA device enforcement

Cobalt Implantable Cardioverter Defibrillators: a. Cobalt VR ICD, Model Numbers: DVPB3D1, DVPB3D4; b. Cobalt DR ICD, Model Numbers: DDPB3D1, DDPB3D4

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Cobalt Implantable Cardioverter Defibrillators: a. Cobalt VR ICD, Model Numbers: DVPB3D1, DVPB3D4...
FDA device enforcement

Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt HF CRT-D, Model Numbers: DTPB2D4, DTPB2D1; b. Cobalt HF Quad CRT-D, Mode...

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a....
FDA device enforcement

Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1, DVPC3D4; b. Crome DR ICD, Model Numbers: DDPC3D1, DDPC3D4

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1, DVPC3D4; ...
FDA device enforcement

Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Crome HF CRT-D, Model Numbers: DTPC2D4, DTPC2D1; b. Crome HF Quad CRT-D, Model N...

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IReported ongoing

Distribution: Worldwide Distribution

Official source for Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. ...
FDA device enforcement

The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device...

The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, ...

Official source for The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during...
FDA device enforcement

Cobalt Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Cobalt XT VR: DVPA2D1, DVPA2D4; b) Cobalt VR: DVPB3D1, DVPB3D4; c) Cobalt XT DR: DDP...

Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IIReported ongoing

Distribution: Distributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, ...

Official source for Cobalt Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Coba...
FDA device enforcement

Crome Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Crome VR: DVPC3D1, DVPC3D4; b) Crome DR: DDPC3D1, DDPC3D4; c) Crome HF: DTPC2D4, DTPC...

Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Class IIReported ongoing

Distribution: Distributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, ...

Official source for Crome Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Crome...
FDA device enforcement

Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.

There have been reports of stent migration.

Class IIReported terminated

Distribution: Distribution was nationwide. There was also military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Bahrain, Belgium, Botswana, Bulgari...

Official source for Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.
FDA device enforcement

Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72

The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."

Class IIReported terminated

Distribution: US, Canada, Northern Mariana Islands

Official source for Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72
FDA device enforcement

Medical Device Identification Cards associated with Astra XT DR Implantable pulse generator

The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."

Class IIReported terminated

Distribution: US, Canada, Northern Mariana Islands

Official source for Medical Device Identification Cards associated with Astra XT DR Implantable pulse generator
FDA device enforcement

Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and ca...

The processing of collected episode data may fail due to a software error.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide. There was military/government distribution. The countries of Andorra, Australia, Austria, Belgium, Brunei Darussalam, Canada, Czech Repub...

Official source for Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following ...

Reasons in these listed records

  • There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
  • The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.
  • Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
  • Devices may lose functionality due to susceptibility to moisture ingress.

These examples come from this result page and are not a ranking of every recall reason.

Checking Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Is there a recall on medtronic inc., cardiac rhythm and heart failure (crhf) right now?

This archive contains 134 records, including 130 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking