The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedu... — FDA record Z-1206-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 20, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a catheter dilator to facilitate deflectable catheter passage, and a guide catheter slitter to remove the deflectable catheter. The deflectable catheter is designed for placement of transvenous devices in or near the bundle of His. It features a deflecting distal section, controlled by the deflectable catheter handle, and an out-of plane curve on the distal tip. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.
Reason
The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.
Classification
Class II
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
1385 devices
Agency-reported status
Ongoing
Source record identifier
Z-1206-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Number: C304-HIS; GTIN: 00763000147013; Lot Numbers: 0010948864, 0010953298, 0010958150, 0010961222, 0010971148, 0011017539, 0011020263, 0011022874, 0011025202, 0011028574, 0011031177

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Belgium, Denmark, France, Germany, Ireland, Italy, Netherlands, Poland, Spain, United Arab Emirates, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 8, 2022
Recall initiation date
Apr 20, 2022
Center classification date
May 27, 2022