Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recall records

130 reported ongoing

134 records in this archive. Latest source date May 10, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVFB2D1 VISIA AF MRI XT OUS DF1, Model Number DVFB2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVFB2D1 VISIA AF MRI XT OUS DF1, Model Number DVFB2D1; Implantable Cardioverter Defibrilla...
FDA device enforcement

ICD-VR VISIA MRI AF XT OUS DF4, Model Number DVFB2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA MRI AF XT OUS DF4, Model Number DVFB2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibril...
FDA device enforcement

ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMB1D4 EVERA MRI XT DF4 US, Model Number DVMB1D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMB1D4 EVERA MRI XT DF4 US, Model Number DVMB1D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMB2D1 EVERA MRI XT OUS DF1, Model Number DVMB2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMB2D1 EVERA MRI XT OUS DF1, Model Number DVMB2D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMB2D4 EVERA MRI XT DF4 INTL, Model Number DVMB2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMB2D4 EVERA MRI XT DF4 INTL, Model Number DVMB2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMC3D1 EVERA MRI S OUS/US DF1, Model Number DVMC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMC3D1 EVERA MRI S OUS/US DF1, Model Number DVMC3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMC3D4 EVERA MRI S DF4 GLOB, Model Number DVMC3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMC3D4 EVERA MRI S DF4 GLOB, Model Number DVMC3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMD3D1 PRIMO MRI, Model Number DVMD3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMD3D1 PRIMO MRI, Model Number DVMD3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVMD3D4 PRIMO MRI, Model Number DVMD3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVMD3D4 PRIMO MRI, Model Number DVMD3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVME3D1 MIRRO MRI, Model Number DVME3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVME3D1 MIRRO MRI, Model Number DVME3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD COBALT XT VR MRI IS1 DF1, Model Number DVPA2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD COBALT XT VR MRI IS1 DF1, Model Number DVPA2D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD COBALT VR MRI IS1 DF1, Model Number DVPB3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD COBALT VR MRI IS1 DF1, Model Number DVPB3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD CROME VR MRI IS1 DF1, Model Number DVPC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD CROME VR MRI IS1 DF1, Model Number DVPC3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD CROME VR MRI DF4, Model Number DVPC3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD CROME VR MRI DF4, Model Number DVPC3D4; Implantable Cardioverter Defibrillators

Reasons in these listed records

  • There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, inc...

These examples come from this result page and are not a ranking of every recall reason.

Checking Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Is there a recall on medtronic inc., cardiac rhythm and heart failure (crhf) right now?

This archive contains 134 records, including 130 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking