Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recall records

130 reported ongoing

134 records in this archive. Latest source date May 10, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

CRTD COBALT HF QUAD MRI IS4 DF4, Model Number DTPB2QQ ; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for CRTD COBALT HF QUAD MRI IS4 DF4, Model Number DTPB2QQ ; Implantable Cardioverter Defibrillators
FDA device enforcement

CRTD CROME HF MRI IS1 DF1, Model Number DTPC2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for CRTD CROME HF MRI IS1 DF1, Model Number DTPC2D1; Implantable Cardioverter Defibrillators
FDA device enforcement

CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

CRTD CROME HF QUAD MRI IS4 DF1, Model Number DTPC2Q1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for CRTD CROME HF QUAD MRI IS4 DF1, Model Number DTPC2Q1; Implantable Cardioverter Defibrillators
FDA device enforcement

CRTD CROME HF QUAD MRI IS4 DF4, Model Number DTPC2QQ; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for CRTD CROME HF QUAD MRI IS4 DF4, Model Number DTPC2QQ; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVAB1D4 VISIA AF US DF4, Model Number DVAB1D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVAB1D4 VISIA AF US DF4, Model Number DVAB1D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR VISIA AF XT OUS DF4, Model Number DDVAB2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA AF XT OUS DF4, Model Number DDVAB2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBB2D1 EVERA XT IS1/DF1 INTL, Model Number DVBB2D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBB2D1 EVERA XT IS1/DF1 INTL, Model Number DVBB2D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBB2D4 EVERA XT DF4 INTL, Model Number DVBB2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBB2D4 EVERA XT DF4 INTL, Model Number DVBB2D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVBC3D4 EVERA S DF4 GLOB, Model Number DVBC3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVBC3D4 EVERA S DF4 GLOB, Model Number DVBC3D4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVEX3E4 EV ICD EV4 PIVOTAL, Model Number DVEX3E4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVEX3E4 EV ICD EV4 PIVOTAL, Model Number DVEX3E4; Implantable Cardioverter Defibrillators
FDA device enforcement

ICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glass...

Class IReported ongoing

Distribution: US Nationwide - Worldwide Distribution

Official source for ICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators

Reasons in these listed records

  • There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, inc...

These examples come from this result page and are not a ranking of every recall reason.

Checking Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Is there a recall on medtronic inc., cardiac rhythm and heart failure (crhf) right now?

This archive contains 134 records, including 130 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking