ICU Medical, Inc. recall records

96 reported ongoing

96 records in this archive. Latest source date Aug 20, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

ICU Medical Oncology Kit w/2 ChemoLock", 13mm Closed Vial Spike w/ChemoLock" Port, ChemoLock" Bag Spike, Ext Set w/ChemoLock" Port, Clamp, Graduated Adapter REF: CL4188

Port weld of drug transfer device may separate or break during use and potentially result in a leak.

Class IIReported ongoing

Distribution: Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY an...

Official source for ICU Medical Oncology Kit w/2 ChemoLock", 13mm Closed Vial Spike w/ChemoLock" Port, ChemoLock" Bag...
FDA device enforcement

ICU Medical ChemosafeLock Vial Adapter REFs: KL-VA001U3 KL-VA002U3 KL-VA131U3 KL-VA201U3 KL-VA202U3 KL-VA321U3

Port weld of drug transfer device may separate or break during use and potentially result in a leak.

Class IIReported ongoing

Distribution: Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY an...

Official source for ICU Medical ChemosafeLock Vial Adapter REFs: KL-VA001U3 KL-VA002U3 KL-VA131U3 KL-VA201U3 KL-VA202...
FDA device enforcement

Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper i...
FDA device enforcement

Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7002-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper ...
FDA device enforcement

Smiths Medical CADD" Yellow Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included 100mL, Product Code 21-7300-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical CADD" Yellow Medication Cassette Reservoir with flow stop, clamp and female Luer. ...
FDA device enforcement

Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-ven...
FDA device enforcement

Smiths Medical CADD Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7302-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical CADD Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vent...
FDA device enforcement

Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubing, clamp and female NRFit connector. Yellow stopper inclu...

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow sto...
FDA device enforcement

SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000

Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS,...

Official source for SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-c...
FDA device enforcement

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2

Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.

Official source for LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2
FDA device enforcement

ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Man...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Official source for ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump
FDA device enforcement

ICU Medical Plum Solos, Item number 400011001; infusion pump

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Man...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Official source for ICU Medical Plum Solos, Item number 400011001; infusion pump
FDA drug enforcement

POTASSIUM CHLORIDE Inj., 20 mEq total in 100 mL flexible container 24 x case, 200 mEq/L, Rx Only, ICU Medical, Inc., Lake Forest, Illinois, 60045, USA, NDC# 0990-7075-26; Bar Co...

Labeling: Label Error on Declared Strength. Cases labeled POTASSIUM CHLORIDE 20 mEq, may contain flexible containers with overwrap mislabeled as 10 mEq. The correct dosage strength of 20 mEq is printed on the labeling affixed to the product flexible container.

Class IReported ongoing

Distribution: Nationwide in the USA

Official source for POTASSIUM CHLORIDE Inj., 20 mEq total in 100 mL flexible container 24 x case, 200 mEq/L, Rx Only,...
FDA drug enforcement

POTASSIUM CHLORIDE Inj., 10 mEq total in 100 mL, 100 mEq/L flexible container, Rx Only, ICU Medical, Inc., Lake Forest, Illinois, 60045, USA, NDC# 0990-7074-26; Bar Code (01)003...

Labeling: Label Error on Declared Strength. Overwrap labeled as Potassium Chloride Inj 10 mEq may contain flexible containers of Potassium Chloride Inj 20 mEq

Class IReported ongoing

Distribution: Nationwide in the USA

Official source for POTASSIUM CHLORIDE Inj., 10 mEq total in 100 mL, 100 mEq/L flexible container, Rx Only, ICU Medic...
FDA device enforcement

ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be ...

Due to finished goods incorrectly packaged and labeled.

Class IIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution in the state of IL and the country of Canada.

Official source for ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the tra...
FDA device enforcement

Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infus...

ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have...

Class IReported ongoing

Distribution: Worldwide distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries of: Argentina, Brazil, Canada, Chile, Colombia, France, Greece, Hong Kong, Italy, Irela...

Official source for Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Syste...
FDA device enforcement

Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following ...

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Class IReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Ara...

Official source for Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010....

Reasons in these listed records

  • Port weld of drug transfer device may separate or break during use and potentially result in a leak.
  • Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
  • ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
  • Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.

These examples come from this result page and are not a ranking of every recall reason.

Checking ICU Medical, Inc.

Is there a recall on icu medical, inc. right now?

This archive contains 96 records, including 96 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking