LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2 — FDA record Z-1763-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 2, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2
Reason
Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
Classification
Class II
Recalling firm
ICU Medical, Inc.
Quantity
12 US, 3 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1763-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLifeShield Infusion Safety Software Suite v2.1 and v2.2 UDI-DI 00887709122802

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 21, 2025
Recall initiation date
May 2, 2025
Center classification date
May 14, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.