Records and reported status

FDA device enforcement

GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
FDA device enforcement

GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed
FDA device enforcement

GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
FDA device enforcement

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
FDA device enforcement

GE Healthcare Revolution CT, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
FDA device enforcement

GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the g...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
FDA device enforcement

Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limi...

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechi...

Official source for Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray...
FDA device enforcement

Discovery XR656HD. X-Ray imaging system.

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limi...

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechi...

Official source for Discovery XR656HD. X-Ray imaging system.
FDA device enforcement

Discovery XR656 HD. X-Ray imaging system.

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limi...

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechi...

Official source for Discovery XR656 HD. X-Ray imaging system.
FDA device enforcement

Optima XR646 HD. X-Ray imaging system.

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limi...

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechi...

Official source for Optima XR646 HD. X-Ray imaging system.
FDA device enforcement

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into ex...

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.

Official source for Portrait Core Services, a set of software services that enable the communication and interaction ...
FDA device enforcement

GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise...

Class IIReported ongoing

Distribution: Worldwide

Official source for GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2...
FDA device enforcement

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise...

Class IIReported ongoing

Distribution: Worldwide

Official source for GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; ...
FDA device enforcement

GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700...

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise...

Class IIReported ongoing

Distribution: Worldwide

Official source for GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; ...
FDA device enforcement

Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 206683...

Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.

Class IIReported ongoing

Distribution: US Nationwide. Global Distribution.

Official source for Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1...
FDA device enforcement

Discovery MR750w 3.0T, whole body magnetic resonance scanner

Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.

Class IIReported ongoing

Distribution: Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.

Official source for Discovery MR750w 3.0T, whole body magnetic resonance scanner

Reasons in these listed records

  • GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is ...
  • The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
  • There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
  • GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the C...

These examples come from this result page and are not a ranking of every recall reason.

Checking GE Medical Systems, LLC

Is there a recall on ge medical systems, llc right now?

This archive contains 124 records, including 124 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking