GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB — FDA record Z-0811-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 25, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
Reason
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Classification
Class II
Recalling firm
GE Medical Systems, LLC
Quantity
53 units
Agency-reported status
Ongoing
Source record identifier
Z-0811-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1) 5555000-3: UDI/DI 00840682115278, Serial Numbers: 260665XR220, 478757OPT, 512SDERLM1, 904244MP5, 912466AMX2, 954WRMCDP2; UDI/DI none, Serial Numbers: 10523MOB01, 201418DIG1, 210297NC200M1, 229276M200, 240566P1, 316858AMX200P1, 401OPTI3, 410368WP1, 443643WP1, 503251XR200, 512SDMCP1, 518243OPT2, 603356OPTIMA, 609267OPT, 609267OPT3, 661949DP1, 661949DP2, 718630OPTIMAX, 760379DP, 801299SB1, 804417OPTIMA, 812237P1, 813745OPT1, 856355OPT, 904244MP1, 904244MP2, 904276DP200, 918477OPT1, 956365OPT1, 956632OPT2, 956632OPT4, XR1037005WK4; 2) 5555000-4: UDI/DI 00840682115278, Serial Numbers: 405717P200, 408259AP8, 518583OPT1, 773792XR200, 7982XR0277, 801230XRA, 904399MP19, 989907DP1; UDI/DI none, Serial Numbers: 1553XR0081, 316962240P1, 361572XR220C, 812265M200, A51043204, FEB520100; 3) A0659MB: UDI/DI 00840682115278, Serial Numbers: XR2262CO.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Jan 15, 2025
Recall initiation date
Nov 25, 2024
Center classification date
Jan 7, 2025