Optima XR646 HD. X-Ray imaging system. — FDA record Z-1046-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 28, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Optima XR646 HD. X-Ray imaging system.
Reason
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
Classification
Class II
Recalling firm
GE Medical Systems, LLC
Quantity
3149 (1697 US; 1452 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-1046-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 00195278526892

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 12, 2025
Recall initiation date
Jan 28, 2025
Center classification date
Feb 3, 2025