Covidien recall records

17 reported ongoing

17 records in this archive. Latest source date Jun 23, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm vol...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR,...

Official source for Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005...
FDA device enforcement

Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-...

Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to al...

Class IReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL,...

Official source for Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT...
FDA device enforcement

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R

Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a ...

Class IReported ongoing

Distribution: OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan

Official source for Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
FDA device enforcement

McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx Only

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

Class IReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC,...

Official source for McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx Only
FDA device enforcement

McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx Only

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

Class IReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC,...

Official source for McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx Only
FDA device enforcement

Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILA...

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pos...

Class IReported ongoing

Distribution: US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh,...

Official source for Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600,...
FDA device enforcement

Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1

Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-ste...

Class IIReported ongoing

Distribution: US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA,...

Official source for Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1
FDA device enforcement

Medtronic Invos, Reusable Infant Sensor Adapter Cable for PM7100, REF: PMAC71RIC

manufacturing defects to sensor cable connectors may result in error codes or non-functioning channel

Class IIReported ongoing

Distribution: US: AK AZ CA CT FL GA IA IN LA MA MD MI MN MO NC NY OH OK PA SC TN TX VA WA WI OUS: Australia Austria Belgium Canary Islands Chile Czech Republic Denmark Finland France Germany ...

Official source for Medtronic Invos, Reusable Infant Sensor Adapter Cable for PM7100, REF: PMAC71RIC
FDA device enforcement

Medtronic Invos, Reusable Sensor Cable for PM7100, REF: PMAC71RSC

manufacturing defects to sensor cable connectors may result in error codes or non-functioning channel

Class IIReported ongoing

Distribution: US: AK AZ CA CT FL GA IA IN LA MA MD MI MN MO NC NY OH OK PA SC TN TX VA WA WI OUS: Australia Austria Belgium Canary Islands Chile Czech Republic Denmark Finland France Germany ...

Official source for Medtronic Invos, Reusable Sensor Cable for PM7100, REF: PMAC71RSC
FDA device enforcement

Puritan Bennett Cuff Pressure Manager, REF: 180-03

Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatm...

Class IIReported ongoing

Distribution: US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands

Official source for Puritan Bennett Cuff Pressure Manager, REF: 180-03
FDA device enforcement

McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx Only

Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.

Class IIReported ongoing

Distribution: US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia C...

Official source for McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx Only
FDA device enforcement

McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only

Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.

Class IIReported ongoing

Distribution: US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia C...

Official source for McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
FDA device enforcement

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY...

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to re...

Class IReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA,...

Official source for Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cann...
FDA device enforcement

COVIDIEN Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed REF 86051 7.0 mm I.D. The Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed are indicated for airway management by oral/nasal ...

Due to potential leaks associate with the Oral/Nasal Tracheal Tube Cuffed inflation system.

Class IIReported ongoing

Distribution: U.S.: AK, AZ, CA, FL, GA, IA, IL, IN, KS MD, MI, MS, NC, NE, NJ, NY, OH, OK, PA, SC, TX, UT, VA, WA, and WV O.U.S.: Canada

Official source for COVIDIEN Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed REF 86051 7.0 mm I.D. The Shiley Lo-Pro Or...

Reasons in these listed records

  • Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off al...
  • Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively duri...
  • Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment a...
  • Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

These examples come from this result page and are not a ranking of every recall reason.

Checking Covidien

Is there a recall on covidien right now?

This archive contains 17 records, including 17 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.