Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB5... — FDA record Z-2446-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100
Reason
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
Classification
Class I
Recalling firm
Covidien
Quantity
34,642
Agency-reported status
Ongoing
Source record identifier
Z-2446-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI(GTIN): 4096600/10884521087798 4096600-01/10884521195998 4096600-02/10884521196001 4096600-03/10884521196018 4096600-04/10884521196025 4096600-05/10884521196032 4096600-06/10884521196377 4097100/4097100 4098300/10884521183636 4098300-01/10884521195943 4098300-02/4098300-02 4098300-03/10884521195967 4098300-05/10884521195981 4098300-06/10884521196384 DL4096600/10884521786509 DL4096600-06/10884521786516 DL4097100/DL4097100 DS4097100/DS4097100 All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Montenegro, Namibia, Nepal, New Zealand, Nigeria, Panama, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Viet Nam, South Africa, Austria, Kenya, Latvia, Lebanon, Mongolia, Nicaragua, Thailand, Tunisia, Algeria, Jordan, Libya, Malawi, Qatar, Pakistan, United Arab Emirates, Oman, Peru, Philippines, Bhutan, Iran - Islamic Republic Of, Bolivia, Kuwait, Egypt, Switzerland, Sudan, Norway, Korea - Republic Of, Bahrain, Ethiopia, Paraguay

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 14, 2024
Recall initiation date
Jun 24, 2024
Center classification date
Aug 2, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.