Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R — FDA record Z-1390-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 26, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Reason
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
Classification
Class I
Recalling firm
Covidien
Quantity
780
Agency-reported status
Ongoing
Source record identifier
Z-1390-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI/GTIN: A8845212054401, 20884521205441, 10884521205444. Lot: 202405258X. Expiration: 06/01/2029

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Apr 2, 2025
Recall initiation date
Feb 26, 2025
Center classification date
Mar 25, 2025