Percussionaire, Inc., REF: P5-TEE, Qty: 1, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve — FDA record Z-3207-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Percussionaire, Inc., REF: P5-TEE, Qty: 1, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve
Reason
Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.
Classification
Class I
Recalling firm
Percussionaire Corporation
Quantity
50,460 units
Agency-reported status
Ongoing
Source record identifier
Z-3207-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityNo UDI/ Lots: 240620, 240610, 240418, 240326, 230612, WO07066, WO07010, WO06020, WO05019, WO04884, WO04827, WO04756, WO04714, WO04656, WO04625, WO04596, WO04553, WO04398, WO04356, WO04061, WO03846, WO03748, WO03688

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 9, 2024
Recall initiation date
Aug 21, 2024
Center classification date
Sep 27, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records