Records and reported status

FDA device enforcement

Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use

IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hyp...

Class IReported ongoing

Distribution: Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE

Official source for Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 2...
FDA device enforcement

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VD...

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Class IReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherland...

Official source for VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasit...
FDA device enforcement

Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an ina...

Class IReported ongoing

Distribution: U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U...

Official source for Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ven...
FDA device enforcement

Percussionaire, Inc., REF: P5-TEE, Qty: 1, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve

Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.

Class IReported ongoing

Distribution: US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey

Official source for Percussionaire, Inc., REF: P5-TEE, Qty: 1, Rx Only. IPV (Intrapulmonary Percussive Ventilation) t...
FDA device enforcement

Percussionaire, Inc., REF: P5-TEE-20, Qty: 20, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve

Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.

Class IReported ongoing

Distribution: US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey

Official source for Percussionaire, Inc., REF: P5-TEE-20, Qty: 20, Rx Only. IPV (Intrapulmonary Percussive Ventilatio...
FDA device enforcement

High Frequency Transport Phasitron Breathing Circuit Kit REF: A50605-D, Rx Only, CE 0123

Due to an assembly error within the breathing circuit, their pressure output of the breathing circuit kit can be affected and have severe impact to the end user

Class IReported ongoing

Distribution: US:AL AR CA CT FL GA IL IN KS LA MI MN MO NC NJ OH OR TN TX UT VA OUS: None

Official source for High Frequency Transport Phasitron Breathing Circuit Kit REF: A50605-D, Rx Only, CE 0123

Reasons in these listed records

  • IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, update...
  • Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
  • As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved ...
  • Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.

These examples come from this result page and are not a ranking of every recall reason.

Checking Percussionaire Corporation

Is there a recall on percussionaire corporation right now?

This archive contains 7 records, including 7 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking