INZONE DETACHMENT SYSTEM, REF: M00345100950 — FDA record Z-2552-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 6, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
INZONE DETACHMENT SYSTEM, REF: M00345100950
Reason
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
Classification
Class II
Recalling firm
Stryker Neurovascular
Quantity
44,937
Agency-reported status
Ongoing
Source record identifier
Z-2552-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 1, 2026
Recall initiation date
Apr 6, 2026
Center classification date
Jun 24, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records