FDA device enforcement

INZONE DETACHMENT SYSTEM, REF: M00345100950

Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical interventi...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC,...

Official source for INZONE DETACHMENT SYSTEM, REF: M00345100950
FDA device enforcement

The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The...

Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.

Class IIReported ongoing

Distribution: U.S. Nationwide distribution in the states of CA and VA.

Official source for The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter wit...
FDA device enforcement

Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x ...

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183...
FDA device enforcement

Trevo ProVue, Catalog: 90184

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Trevo ProVue, Catalog: 90184
FDA device enforcement

Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
FDA device enforcement

Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PAC...

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 914...
FDA device enforcement

Trevo Trak 21 Microcatheter, Catalog: 90338

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Trevo Trak 21 Microcatheter, Catalog: 90338
FDA device enforcement

FlowGate2 8F x 85cm, Catalog: 90485 FlowGate2 8F x 95cm, Catalog: 90495

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for FlowGate2 8F x 85cm, Catalog: 90485 FlowGate2 8F x 95cm, Catalog: 90495
FDA device enforcement

DAC 044 115cm, Catalog: 90760

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for DAC 044 115cm, Catalog: 90760
FDA device enforcement

Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCH...

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high level...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA,...

Official source for Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Ca...
FDA device enforcement

The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Sync...

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain ...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ,...

Official source for The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside d...
FDA device enforcement

The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewire...

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain ...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ,...

Official source for The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are avail...
FDA device enforcement

The Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are ...

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain ...

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ,...

Official source for The Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available i...
FDA device enforcement

The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a luer hub on the proximal end. The cathete...

Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA, MI, MS, SC, IL.

Official source for The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with rad...
FDA device enforcement

Trevo Trak 21 Microcatheter, REF: 90338. TREVO NXT 3X32+TRAK21 2-PACK - CE, REF: 91412; TREVO NXT 4X28+TRAK21 2-PACK - CE, REF: 91413; TREVO NXT 4X41+TRAK21 2-PACK - CE, REF: 91...

Microcatheter product label on the carton is missing the microcatheter product drawing/illustration.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of CA, OR, PA, NY, DC, IL, WA, NH, NJ, OK, FL, AZ, AL, GA, TX, WI, VA and the countries of DE, IT, AT.

Official source for Trevo Trak 21 Microcatheter, REF: 90338. TREVO NXT 3X32+TRAK21 2-PACK - CE, REF: 91412; TREVO NXT...

Reasons in these listed records

  • Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure ...
  • Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.
  • Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may ...
  • Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro ...

These examples come from this result page and are not a ranking of every recall reason.

Checking Stryker Neurovascular

Is there a recall on stryker neurovascular right now?

This archive contains 15 records, including 15 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking