Abbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only — FDA record Z-2430-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 22, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only
Reason
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
Classification
Class II
Recalling firm
Abbott Medical
Quantity
5,900 units
Agency-reported status
Ongoing
Source record identifier
Z-2430-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Serial numbers/UDI: 05415067020253

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Turkey, United Arab Emirates, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 31, 2024
Recall initiation date
May 22, 2024
Center classification date
Jul 25, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records