FDA device enforcement

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

Class IIReported ongoing

Distribution: US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH,...

Official source for Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
FDA device enforcement

Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
FDA device enforcement

Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of...

Class IIReported ongoing

Distribution: Worldwide distribution.

Official source for Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
FDA device enforcement

Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports ...

Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

Class IIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution in the states of FL, SC, and WV and the country of Spain. There were five (5) LSP202V devices distributed within the United States (3 in...

Official source for Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF L...
FDA device enforcement

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 softwar...

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (

Class IIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, N...

Official source for The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, ded...
FDA device enforcement

Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Class IIReported ongoing

Distribution: New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Official source for Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile
FDA device enforcement

Abbott Navitor Transcatheter Aortic Heart Valve, 29 mm, NVTR-29, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Class IIReported ongoing

Distribution: New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Official source for Abbott Navitor Transcatheter Aortic Heart Valve, 29 mm, NVTR-29, Sterile
FDA device enforcement

Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Class IIReported ongoing

Distribution: New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Official source for Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile
FDA device enforcement

Abbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Class IIReported ongoing

Distribution: New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Official source for Abbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, Sterile
FDA device enforcement

Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Class IIReported ongoing

Distribution: New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Official source for Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
FDA device enforcement

Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted heart device to Merlin.net" Patient Care N...

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Class IIIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA...

Official source for Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer sto...
FDA device enforcement

Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand" (MoD) capability is a transmitter configured for multiple patient use....

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Class IIIReported ongoing

Distribution: Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA...

Official source for Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the Merlin...
FDA device enforcement

Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA,...

Official source for Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only
FDA device enforcement

Abbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA,...

Official source for Abbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only
FDA device enforcement

Abbott Infinity 7, Implantable Pulse Generator, REF: 6662, SterileEO, Rx Only

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA,...

Official source for Abbott Infinity 7, Implantable Pulse Generator, REF: 6662, SterileEO, Rx Only
FDA device enforcement

Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA,...

Official source for Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only
FDA device enforcement

Abbott Liberta RC, Implantable Pulse Generator, REF: 62400, STERILE, RxONLY

Deep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will recur approximately every 50 days thereafter. This unexpected lost of stimulation may result in loss of therapy.

Class IIReported ongoing

Distribution: US: Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NE, NH, NJ, NY, OH OK, OR, PA, SC, SD, TX, UT, WI,...

Official source for Abbott Liberta RC, Implantable Pulse Generator, REF: 62400, STERILE, RxONLY
FDA device enforcement

Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only

The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA,...

Official source for Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only
FDA device enforcement

Abbott Proclaim 5 Implantable Pulse Generator, REF 3661, Sterile EO, Rx Only

The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA,...

Official source for Abbott Proclaim 5 Implantable Pulse Generator, REF 3661, Sterile EO, Rx Only
FDA device enforcement

Abbott Proclaim XR 7 Implantable Pulse Generator, REF 3662, Sterile EO, Rx Only

The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA,...

Official source for Abbott Proclaim XR 7 Implantable Pulse Generator, REF 3662, Sterile EO, Rx Only

Reasons in these listed records

  • Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
  • There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with ...
  • Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.
  • Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (

These examples come from this result page and are not a ranking of every recall reason.

Checking Abbott Medical

Is there a recall on abbott medical right now?

This archive contains 39 records, including 39 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking