MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP — FDA record Z-2373-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 28, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP
Reason
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Classification
Class I
Recalling firm
Covidien LP
Quantity
299 units
Agency-reported status
Ongoing
Source record identifier
Z-2373-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 0884521128010 LOT Numbers: 2133700228

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 23, 2023
Recall initiation date
Jun 28, 2023
Center classification date
Aug 11, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records