Covidien, LP recall records

76 reported ongoing

76 records in this archive. Latest source date Jun 24, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0 CLR 15CM P12X12 VLOCM0003 VLOCM0013 V-...

Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.

Class IIReported ongoing

Distribution: US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finl...

Official source for V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VL...
FDA device enforcement

V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON BOX01056V2 BOX BYPASS DR LAGAE BOX01482V2 BOX BY...

Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.

Class IIReported ongoing

Distribution: US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finl...

Official source for V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FIN...
FDA device enforcement

Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31

Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.

Class IIReported ongoing

Distribution: US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finl...

Official source for Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C1...
FDA device enforcement

Surgipro SP-623 SURGIPRO* 2-0 BLU 75CM SC2 X36 SP-623

Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.

Class IIReported ongoing

Distribution: US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finl...

Official source for Surgipro SP-623 SURGIPRO* 2-0 BLU 75CM SC2 X36 SP-623
FDA device enforcement

Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX

Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.

Class IIReported ongoing

Distribution: US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finl...

Official source for Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
FDA device enforcement

Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;

The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.

Class IIReported ongoing

Distribution: Domestic: Nationwide distribution; International: Austria, Canada, Denmark, France, Italy;

Official source for Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
FDA device enforcement

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, ...

Official source for Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45...
FDA device enforcement

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, ...

Official source for Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMI...
FDA device enforcement

Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm x 75cm (CFN CH07-40-75US), 8mm x 40mm x...

Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

Class IIReported ongoing

Distribution: Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Canada, Canary Islands, China, France, French Guiana, Germany, Greece, Guadeloupe, India, Ire...

Official source for Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40...
FDA device enforcement

Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of a...

Potential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the state of NJ and the countries of Australia, Japan, New Zealand, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand.

Official source for Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pedia...
FDA device enforcement

Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute ...

Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, ...

Official source for Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip,...
FDA device enforcement

Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, ap...

Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, ...

Official source for Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin ...
FDA device enforcement

Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB

The potential for trocar seal disengagement when using mesh products incorrectly with the device.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Cz...

Official source for Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB
FDA device enforcement

Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL

The potential for trocar seal disengagement when using mesh products incorrectly with the device.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Cz...

Official source for Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL
FDA device enforcement

Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL

The potential for trocar seal disengagement when using mesh products incorrectly with the device.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Cz...

Official source for Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL
FDA device enforcement

Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BTNL

The potential for trocar seal disengagement when using mesh products incorrectly with the device.

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Cz...

Official source for Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BTNL
FDA device enforcement

Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled arti...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

Official source for Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Reasons in these listed records

  • Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
  • The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
  • A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
  • Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

These examples come from this result page and are not a ranking of every recall reason.

Checking Covidien, LP

Is there a recall on covidien, lp right now?

This archive contains 76 records, including 76 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking