Ventilator HAMILTON-C6, PN: 160021 — FDA record Z-2020-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 15, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ventilator HAMILTON-C6, PN: 160021
Reason
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Classification
Class I
Recalling firm
Hamilton Medical AG
Quantity
68
Agency-reported status
Ongoing
Source record identifier
Z-2020-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 07630002808590, Software Versions: SW v1.1.4, SW v1.1.5, SW v1.1.6

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 26, 2024
Recall initiation date
May 15, 2024
Center classification date
Jun 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.