Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen des...

Class IReported ongoing

Distribution: US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH,...

Official source for Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILT...
FDA device enforcement

IntelliCuff, Portable automatic cuff pressure controller REF: 951001

Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.

Class IIReported ongoing

Distribution: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada.

Official source for IntelliCuff, Portable automatic cuff pressure controller REF: 951001
FDA device enforcement

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lea...

Class IReported ongoing

Distribution: US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Official source for Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, u...
FDA device enforcement

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and...

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interr...

Class IReported ongoing

Distribution: US Nationwide distribution.

Official source for Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ve...
FDA device enforcement

Ventilator HAMILTON-C6, PN: 160021

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error in...

Class IReported ongoing

Distribution: US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.

Official source for Ventilator HAMILTON-C6, PN: 160021
FDA device enforcement

HAMILTON-C6, REF: 160021

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, I...

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

Official source for HAMILTON-C6, REF: 160021
FDA device enforcement

Hamilton-C6

Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.

Class IIReported ongoing

Distribution: U.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN, KS, KY, MA, MI, MO, NV, NM, NY, NC, OH, PA, Puerto Rico, SC, TX, VA, WA, WV, WI, and WY.

Official source for Hamilton-C6

Reasons in these listed records

  • Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising:...
  • Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
  • Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapn...
  • A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

These examples come from this result page and are not a ranking of every recall reason.

Checking Hamilton Medical AG

Is there a recall on hamilton medical ag right now?

This archive contains 8 records, including 8 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking