Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Ti... — FDA record Z-1596-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 23, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Classification
Class II
Recalling firm
I.T.S. GmbH
Quantity
101 units
Agency-reported status
Ongoing
Source record identifier
Z-1596-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1. Article Number (Lot Numbers): 21311-11 (26/F62573), 21311-15 (26/F65048, 26/F67521), 21311-7 (26/F66165, 26/F67825), 21312-11 (26/F65049, 26/F67826), 21312-15 (26/F66166), 21312-7 (26/F65643, 26/F67827). 2. Article Number (Lot Numbers): 21094-10 (643/032304, 643/082309), 21094-4 (643/0572, 643/0571), 21094-6 (643/112202), 21094-8 (643/012309). 3.Article Number (Lot Numbers): 21093-4 (679/032311, 679/112331), 21093-6 (679/112332).

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 1, 2026
Recall initiation date
Jan 23, 2026
Center classification date
Mar 20, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records