I.T.S. GmbH recall records

17 reported ongoing

17 records in this archive. Latest source date Jan 23, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II...
FDA device enforcement

Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Claviculaplate with Angular Stability, with the below descriptions and associated article numbers...
FDA device enforcement

Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, ...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal H...
FDA device enforcement

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Ar...
FDA device enforcement

Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 213...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers....
FDA device enforcement

Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olec...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Olecranonplate with Angular Stability with the below descriptions and corresponding article numbe...
FDA device enforcement

I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forea...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding articl...
FDA device enforcement

I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 3140...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous ...
FDA device enforcement

Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers....
FDA device enforcement

I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 216...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding ...
FDA device enforcement

Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, ...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Distal Humeral Plates with Angular Stability with the below descriptions and corresponding articl...
FDA device enforcement

I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 2130...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article ...
FDA device enforcement

I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below de...
FDA device enforcement

FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Can...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descript...
FDA device enforcement

HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26,...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers:...
FDA device enforcement

HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1...
FDA device enforcement

Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: ...

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class IIReported ongoing

Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official source for Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below description...

Reasons in these listed records

  • Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

These examples come from this result page and are not a ranking of every recall reason.

Checking I.T.S. GmbH

Is there a recall on i.t.s. gmbh right now?

This archive contains 17 records, including 17 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking