Product
cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190
Reason
Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
Classification
Class II
Recalling firm
Roche Molecular Systems, Inc.
Quantity
14 units
Agency-reported status
Ongoing
Source record identifier
Z-0828-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 6, 2022
Recall initiation date
Nov 16, 2021
Center classification date
Mar 28, 2022