Roche Molecular Systems, Inc. recall records

4 reported ongoing

5 records in this archive. Latest source date Dec 21, 2022.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

Records and reported status

FDA device enforcement

cobas SARS-CoV-2 & Influenza A/B Qualitative assay for use on the cobas 6800/8800 Systems; Material number 09233474190

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influe...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. Foreign distribution worldwide.

Official source for cobas SARS-CoV-2 & Influenza A/B Qualitative assay for use on the cobas 6800/8800 Systems; Materi...
FDA device enforcement

cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influe...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. Foreign distribution worldwide.

Official source for cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/880...
FDA device enforcement

cobas Influenza A/B & RSV UC (Utility Channel) Qualitative nucleic acid test for use on the cobas 6800/8800 Systems; Material number 09233962190

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influe...

Class IIReported ongoing

Distribution: Domestic distribution nationwide. Foreign distribution worldwide.

Official source for cobas Influenza A/B & RSV UC (Utility Channel) Qualitative nucleic acid test for use on the cobas...
FDA device enforcement

cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) in laboratory settings.

Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.

Class IIReported terminated

Distribution: International distribution in the countries of Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Italy, Lithuania, Poland, Portugal, Spain, Sweden, Switzerland...

Official source for cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PC...
FDA device enforcement

cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190

Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influe...

Class IIReported ongoing

Distribution: Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.

Official source for cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190

Reasons in these listed records

  • The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true ...
  • Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.
  • Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.

These examples come from this result page and are not a ranking of every recall reason.

Checking Roche Molecular Systems, Inc.

Is there a recall on roche molecular systems, inc. right now?

This archive contains 5 records, including 4 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. Use the label-check guide if you need help locating an identifier.

Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. Check source coverage and the agency's current notices.

Where can I find older records?

Use monthly archives or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

Keep checking