Recall archive / FDA device enforcement
Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L,... — FDA record Z-0201-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 28, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L, VLOCN304L, VLOCN306L, VLOCN308L
- Reason
- Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.
- Classification
- Class II
- Recalling firm
- Covidien, LP
- Quantity
- 30658
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0201-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model VLOCN004L UDI-DI: 20884521072579 Lot N0B0834Y Model VLOCN006L UDI-DI: 20884521068824 Lots N0B0660Y N0J0042Y N0J0896Y N1A0931Y N1C0644Y N1L0438Y N2E0205Y N2E0350Y Model VLOCN008L UDI-DI: 20884521073606 Lots N0B0787Y N0E0302Y N1C0706Y N1G0856Y Model VLOCN204L UDI-DI: 20884521072609 Lots N0B0001Y N0E0552Y N0G0568Y N0J0894Y Model VLOCN206L UDI-DI: 20884521068831 Lots N0B0957Y N0G0569Y N0J0933Y N0L0198Y N0M0714Y N1A0932Y N1C0558Y N1C0770Y N1F0047Y N1F0796Y N1G0758Y N1K0184Y N2E0270Y N2E0298Y N9L0954Y Model VLOCN208L UDI-DI: 20884521073774 Lots N0B0772Y N0E0990Y N0G0570Y N0J0870Y N1C0557Y N1F0109Y N1F0132Y N1G0101Y N1J0032Y Model VLOCN304L UDI-DI: 20884521072616 Lots N0B0833Y N0G0487Y Model VLOCN306L UDI-DI: 20884521072654 Lots N0B0851Y N0E0558Y N0G0719Y Model VLOCN308L UDI-DI: 20884521073811 Lot N0G0324Y |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Canada Canary Islands Chile Croatia Cyprus Denmark Egypt Finland France Germany Greece Hungary Ireland Israel Italy Jordan Kazakhstan Kuwait Lithuania Luxembourg Mexico Netherlands New Zealand Norway Poland Portugal Puerto Rico Qatar Reunion Romania Russian Federation Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 23, 2022
- Recall initiation date
- Sep 28, 2022
- Center classification date
- Nov 11, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.