Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L,... — FDA record Z-0201-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 28, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L, VLOCN304L, VLOCN306L, VLOCN308L
Reason
Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.
Classification
Class II
Recalling firm
Covidien, LP
Quantity
30658
Agency-reported status
Ongoing
Source record identifier
Z-0201-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel VLOCN004L UDI-DI: 20884521072579 Lot N0B0834Y Model VLOCN006L UDI-DI: 20884521068824 Lots N0B0660Y N0J0042Y N0J0896Y N1A0931Y N1C0644Y N1L0438Y N2E0205Y N2E0350Y Model VLOCN008L UDI-DI: 20884521073606 Lots N0B0787Y N0E0302Y N1C0706Y N1G0856Y Model VLOCN204L UDI-DI: 20884521072609 Lots N0B0001Y N0E0552Y N0G0568Y N0J0894Y Model VLOCN206L UDI-DI: 20884521068831 Lots N0B0957Y N0G0569Y N0J0933Y N0L0198Y N0M0714Y N1A0932Y N1C0558Y N1C0770Y N1F0047Y N1F0796Y N1G0758Y N1K0184Y N2E0270Y N2E0298Y N9L0954Y Model VLOCN208L UDI-DI: 20884521073774 Lots N0B0772Y N0E0990Y N0G0570Y N0J0870Y N1C0557Y N1F0109Y N1F0132Y N1G0101Y N1J0032Y Model VLOCN304L UDI-DI: 20884521072616 Lots N0B0833Y N0G0487Y Model VLOCN306L UDI-DI: 20884521072654 Lots N0B0851Y N0E0558Y N0G0719Y Model VLOCN308L UDI-DI: 20884521073811 Lot N0G0324Y

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Canada Canary Islands Chile Croatia Cyprus Denmark Egypt Finland France Germany Greece Hungary Ireland Israel Italy Jordan Kazakhstan Kuwait Lithuania Luxembourg Mexico Netherlands New Zealand Norway Poland Portugal Puerto Rico Qatar Reunion Romania Russian Federation Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 23, 2022
Recall initiation date
Sep 28, 2022
Center classification date
Nov 11, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.