Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the follo... — FDA record Z-0139-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 6, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.
Reason
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
Classification
Class II
Recalling firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Quantity
13
Agency-reported status
Ongoing
Source record identifier
Z-0139-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 30, 2024
Recall initiation date
Sep 6, 2024
Center classification date
Oct 22, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.